GC工作过程 Gas Chromatography (Go] is based on the ability of a chromatography column, containing the stationary phase, to separate the components of the sample as it is carried through the column by a flow of inert carrier gas. Only volatile samples can be analysed, and elevated temperatures are used to ensure adequate vapour pressures of each sample component. GC can be used to determine the number of components in a sample, and the concentration and purity of each component. The block diagram below shows the main stages of a GC analysis. Prepare Inject onto eparate Detect dentify and Sample Column Components Sample Quantify Components
11
气相色谱仪 5 12 I0 图192气相色谱流程示意图 1—高压气瓶(载气源)2一减压阀 3一气流调节阀 4—净化干燥管 5-压力表 6一热导池 7-进样口 8—色谱柱 9—恒温箱(虚线内) 10—皂膜流量计 11一测量电桥12—记录仪
12 气相色谱仪
4.2气相色谱仪简介 1、检测器的分类 气相色谱检测器是把载气里被分离的各组分的 浓度或质量转换成电信号的装置。目前检测器的种 类多达数十种。 (1)根据检测原理的不同,可将其分为浓度型检测器和 质量型检测器。 浓度型检测器:测量的是载气中某组分浓度瞬间的变化, 即检测器的响应值和组分的浓度成正比。如热导检测器和电 子捕获检测器。 13
13 1、 检测器的分类 气相色谱检测器是把载气里被分离的各组分的 浓度或质量转换成电信号的装置。目前检测器的种 类多达数十种。 (1)根据检测原理的不同,可将其分为浓度型检测器和 质量型检测器。 浓度型检测器: 测量的是载气中某组分浓度瞬间的变化, 即检测器的响应值和组分的浓度成正比。如热导检测器和电 子捕获检测器。 4.2 气相色谱仪简介
4.2气相色谱仪简介 质量型检测器:测量的是载气中某组分进入检测器的速度 变化,即检测器的响应值和单位时间内进入检测器某组分的 量成正比。如火焰离子化检测器和火焰光度检测器等。 (2)按照检测化合物的范围和种类可分为通用型、选择 性检测器 通用型检测器: 热导池检测器(TCD)、氢火焰离子化检测器(FID) 选择性检测器: 电子捕获检测器ECD),测含电负性的物质; 火焰光度检测器(FPD),测含硫磷的化合物; 14 氮磷检测器NPD,测合氮磷的化合物
14 质量型检测器:测量的是载气中某组分进入检测器的速度 变化,即检测器的响应值和单位时间内进入检测器某组分的 量成正比。如火焰离子化检测器和火焰光度检测器等。 4.2 气相色谱仪简介 (2)按照检测化合物的范围和种类可分为通用型、选择 性检测器。 通用型检测器: 热导池检测器(TCD)、氢火焰离子化检测器(FID)。 选择性检测器: 电子捕获检测器(ECD),测含电负性的物质; 火焰光度检测器(FPD),测含硫磷的化合物; 氮磷检测器(NPD),测含氮磷的化合物
2、常用的检测器: 熟导检测器(TCD)、火焰禽子化检测器(FID)、电 子捕获检测器(ECD)。 (1)热导检测器是根据不同的物质具有不同的热导系数原 理制成的。 由于结构简单,性能稳定,几乎对所有物质都有响应, 通用性好,而且线性范围宽,价格便宜,因此是应用最广, 最成熟的一种检测器。其主要缺点是灵敏度较低。 15
15 2、 常用的检测器: 热导检测器(TCD)、火焰离子化检测器(FID)、电 子捕获检测器(ECD)。 (1)热导检测器是根据不同的物质具有不同的热导系数原 理制成的。 由于结构简单,性能稳定,几乎对所有物质都有响应, 通用性好,而且线性范围宽,价格便宜,因此是应用最广, 最成熟的一种检测器。其主要缺点是灵敏度较低